If the Emergency Off switch has been used to shutdown the Echo Liquid Handler, use the following procedure to start up again:
1. Determine and resolve the original problem.
2. Turn the Echo Liquid Handler back on by twisting the Emergency Off switch clockwise. It will pop out to its normal
operating position.
When the Emergency Off switch is returned to its normal position, power is then supplied to most internal systems and most
internal mechanisms are enabled. The Echo Liquid Handler stage motors are not automatically enabled. The operator must
explicitly command a stage homing procedure. For more information , see Motion Calibration.
Note: If you cannot determine the original problem, or you cannot correct the problem, you must arrange for a Labcyte
field engineer to restore your Echo Liquid Handler to normal use.