EasyManua.ls Logo

laerdal SimNewB - Blood Pressure and Pulse Parameters

laerdal SimNewB
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
27
ENGLISH
INTRODUCTION SETUP GUIDE
CLINICAL
SIMULATION PC UTILITIES
DETAIL
SPECIFICATIONS
TROUBLE-
SHOOTING
REMOTE
CONTROL
Blood Pressure and Pulse parameters
On the main display, use the navigation keys to select the Blood
Pressure (BP) functional area (Figure 13), which shows the cur-
rent BP and the manikin’s pulse status (on / off).
Use the
BP- and BP+ quick execution keys to change both sys-
tolic and diastolic blood pressures in steps of 2, their original
difference being maintained.
Pressing the Shift-key reveals the alternative quick execution
functions (Figure 14). Selecting
Pulse will toggle the manikin’s
pulse function on / off, as indicated in the BP functional area:
P: On or P: Off.
Set Blood Pressure menu
When the BP functional area is selected, press OK to open the
Set Blood Pressure menu (Figure 15). Use the left / right navi-
gation keys to select the parameter you wish to change, then
use the up / down navigation keys or the numeric keys to ad-
just each setting.
Using the
BP- and BP+ quick execution keys
- When
Systolic is selected (as in Figure 15), only systolic will
change, but diastolic will be kept at least 10 mmHg below
systolic.
- When
Diastolic is selected, only diastolic will change, but sys-
tolic will be kept at least 10 mmHg above diastolic.
- When
Vol, Gap or Pulse is selected, both systolic and diastolic
will change, and will maintain the difference, each time the
BP- or BP+ quick execution keys, are used.
Set PEA quick execution key will set BP to 0/0 and breathing
rate to 0.
Reset PEA will set the parameters back to what they
were before Set PEA was selected. Increasing the BP or
breathing rate from a PEA state in the main display will auto-
matically reset PEA.
If the running rhythm is a rhythm with no BP, i.e. VF or Asystole,
BP will automatically be set to 0/0. The BP is not automatically
set back to normal if a perfusing rhythm is chosen.
Use the or quick execution keys to alter volu-
me of Korotkoff sounds directly.
Auscultatory
Gap (Korotkoff sound disappears in part
of Phase II) can be toggled on / off using the up / down naviga-
tion keys.
Press OK to activate selections.
Figure 13
Figure 14
Figure 15
SimNewB.qxp:SimNewB DFU EN 3/25/08 4:24 PM Page 27

Other manuals for laerdal SimNewB

Related product manuals