Logging of simulation events
On the main display, use the navigation keys to select the Log
functional area (Figure 21).
The SimNewB™ Neonatal Simulator automatically generates a
log of events and parameter changes from the moment the
system is turned on. Elapsed time, logging status and name for
the current log is displayed in the upper left corner of the dis-
play. Unless saved, the log will be erased when the Control
Unit is turned off.
The
Pause function (also quick execution function
stops the log clock. Any events or changes in settings are not
logged. The
Resume function (also quick execution function )
resumes logging and clock. The
Restart function resets
the log timer and erases the content of the current log.
When the log functional area is selected, press OK to enter the
log control menu (Figure 22) or select functions directly from
quick execution functions. To return to the main menu, press
OK (activate changes) or C (no change).
Save Log
The
Save quick execution function pauses the log and opens
the Save Log field. The log is given a default name by the
SimNewB system, LOGxx, where xx is a sequential number.
This name can be modified, using the backspace quick
execution function and the alphanumeric keys. Press OK to save
the log and return to main display, and press to resume the
log.
Up to 25 logs can be saved in the Control Unit. Logs are er-
ased when transferred to a PC, or they can be erased using the
Clear All Logs function.
Saved logs can be retrieved later using the SimNewB system
PC utility program. Logs can be transferred to the PC for per-
manent storage, additional comments and printing.
Log Events
The
Event quick execution function opens the Log Event menu
(Figure 23), from which you can enter simulation events into
the log. The events are chosen from a list of up to 14 that are
stored in the Remote Control. To log an event, select the event
using the navigation keys, and press OK. You can also enter the
numeric identifier directly with the numeric keys.
The first twelve events are preprogrammed. Events can be
deleted using the
Delete quick execution function.
REMOTE
CONTROL
30
Figure 21
Figure 22
Figure 23
SimNewB.qxp:SimNewB DFU EN 3/25/08 4:24 PM Page 30

Other manuals for laerdal SimNewB

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the laerdal SimNewB and is the answer not in the manual?

laerdal SimNewB Specifications

General IconGeneral
Brandlaerdal
ModelSimNewB
CategoryMedical Equipment
LanguageEnglish

Related product manuals