CALSAVE Calibration Save Command
Input: CALSAVE[term]
Remarks: Finalizes the calibration procedure. This command
calculates and stores a checksum of the calibration memory.
This checksum is used on power up to determine if the
instrument is calibrated properly. If this command is
not sent after a calibration, the instrument will display
an "Err 03" on power up. Refer to Paragraph 5.3.1.
CALZ Zero Offset Calibration Constant Command
Input: CALZ <input>, <type>, <value>[term]
Format: n,n,±nnnnnnn
<input> Specifies the input or analog output the zero offset
calibration constant will be provided to. Valid
entries are 1 for input and 2 for the analog output.
<type> Specifies the input sensor type. Valid entries are:
0 = Silicon Diode or Analog Out Voltage Mode
1 = GaAlAs Diode or Analog Out Current Mode
2 = 100 Plat/250
3 = 100 Plat/500
4 = 1000 Plat
5 = NTC RTD
<value> Zero offset calibration constant value.
Remarks: Provides the zero offset calibration constant for the input or
analog output.
CALZ? Zero Offset Calibration Constant Query
Input: CALZ? <input>, <type>[term]
Format: n,n
<input> 1, or 2
<type> 0 – 5
Returned: <value>[term]
Format: ±nnnnnnn (Refer to command for description)