6300 Rev H User’s Manual
125-9106A Page 75
Command Description
Data sent from
PC
Data sent from
6300
SU=n Set the Static units to n, per the following table
n Static unit
1 Feet
2 Meters
3 inHg
4 Mbar
5 Psi
6 mmHg
7 kpa
n 0 if OK; or
Error code
PM Read the current Pitot mode None n, where n is 1 to
3
PM=n Set the Pitot mode to n, per the following
n Pitot mode
1 Measure
2 Leak
3 Control
n 0 if OK; or
Error code
SM Read the current Static mode None n, where n is 1 to
3
SM=n Set the Static mode to n, per the following
n Static mode
1 Measure
2 Leak
3 Control
n 0 if OK; or
Error code
PJ=n Jog the Pitot Target value Up (n=0) or Down (n=1)
n 0 if OK; or
Error code
SJ=n Jog the Static Target value Up (n=0) or Down
(n=1)
n 0 if OK; or
Error code
KJ=n Jog the Pitot Rate Target value Up (n=0) or Down
(n=1)
n 0 if OK; or
Error code
RJ=n Jog the Static Rate Target value Up (n=0) or Down
(n=1)
n 0 if OK; or
Error code
Lm Read the current limit value for Limit m, where m
defines the type of limit per the table below.
m Limit type
1 Max knots
2 Min knots
3 Max knots/min
4 Max Mach
5 Max Feet
6 Min Feet
7 Max Feet/min
None Value for Limit
m