Lighthouse ApexPortable Operating Manual
A-12 248083440-1 Rev 5
30009 unsigned integer Particle Channel 1 [high]
30010 unsigned integer Particle Channel 1 [low]
30011 unsigned integer Particle Channel 2 [high]
30012 unsigned integer Particle Channel 2 [low]
30013 unsigned integer Particle Channel 3 [high]
30014 unsigned integer Particle Channel 3 [low]
30015 unsigned integer Particle Channel 4 [high]
30016 unsigned integer Particle Channel 4 [low]
---
30063 IEEE Float LASER Supply [low]
30064 IEEE Float LASER Supply [high] (percentage based on
LASER Supply Reference Value)
30065 IEEE Float Background Voltage [low] (percentage based
on Calibration Reference value) -
ApexPortable
30066 IEEE Float Background Voltage [high]
---
30069 IEEE Float LASER Voltage [low] (Percentage based on
LASER Reference value)
30070 IEEE Float LASER Voltage [high]
Table A-12 Data Registers
Register Data Type Description
Table A-13: Instrument Current Status
Value Action
0 Service LASER Status
0 = LASER is good. 1 - LASER needs to be serviced.
1 Bad Flow Status
0 = Flow rate is good. 1 = Flow rate is bad.
2 Particle Overflow Status
0 = No overflow 1 = Overflow occurred.