ApexPortable MODBUS Register Map v1.50
248083440-1 Rev 5 A-13
3 Instrument Service Status
0 = Instrument is working correctly.
1 = Service light is on. Instrument malfunction detected.
4 Threshold High Status
0 = Threshold not exceeded. 1 = Threshold exceeded
5 Threshold Low Status
0 = Threshold not exceeded. 1 = Threshold exceeded.
6 Instrument Sampler Status
0 = Nominal Operation. 1 = Sampler Error.
7 ApexPortable:
0 = LASER power in spec. 1 = LASER power out of
spec.
8 ApexPortable:
LASER current status
0 = LASER current in spec. 1 = LASER power out of
spec.
9 ApexPortable:
LASER supply status
0 = LASER supply in spec. 1 = LASER supply out of
spec.
10 ApexPortable:
LASER life status
0 = LASER life in spec. 1 = LASER life out of spec.
11 ApexPortable:
No flow status
0 = There is flow in the unit. 1 = no flow in unit.
12 ApexPortable:
Photoamp supply status
0 = Photoamp supply in spec. 1 = Photoamp supply out
of spec.
13 ApexPortable:
Background status
0 = Photoamp background good. 1 = Photoamp back-
ground out of spec.
Table A-13: Instrument Current Status
Value Action