Operation - Particle Mode
5HY 4-19
PARTICLE Display
Data on the instrument can be displayed in Differential (DIFF) or
Cumulative (CUML) counts.
For example, the cumulative count for the 1m channel is the sum of
that channel’s count + 2.5m count + the 5m count.
The differential count for the 1m channel is the number of particles
between 1m and 2.5m.
The data displays on the MAIN screen according to whichever is
selected (DIFF or CUML).
The data format is either Raw (RAW) or Normalized (NORM). Raw
data displays the actual number of particles counted. Normalized data
shows particle concentrations calculated from the raw data based on the
settings chosen (ft
3
or m
3
).
Volume of Air = Sample time (minutes) x FlowRate (CFM)
Normalized Data = Number of Particles/Volume of Air
Press BACK to return to the Configuration screen or press MAIN to
return to the MAIN screen.
ALARM
The user can enable alarming on specific channels as illustrated in
Figure 4-19.
Figure 4-19 Particle Alarm Configuration, 2 channels enabled
Threshold Button