Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual
4-20 5HY
Note: Alarming is only
applicable for AUTO and
MANUAL mode. It applies
only to Raw particle counts
even if the instrument is
displaying Normalized
data.
If the instrument is set to
Differential data, the alarm
threshold will apply to the
differential counts.
If the instrument is set to
display Cumulative data,
the alarm threshold will
apply to the cumulative
counts.
To enable the alarming for any channel, press the “X” next to that
channel. When a checkmark is displayed, that channel is enabled for
alarming. Press the checkmark to disable the alarming for that channel.
Alarm Threshold
Press the threshold button next to the enabled channel in order to set the
alarm threshold for that channel. See Figure 4-20.
Figure 4-20 Configure Alarm Threshold
Enter the desired alarm threshold for the selected particle channel, in
number of particles, then press ENTER. The threshold value will be
updated on the Particle Alarm screen as shown in Figure 4-21.