Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual
4-46 5HY
PRINT RANGE (BUFFER REPORT)
The instrument stores up to 3000 records in its data buffer. The Buffer
Report prints all or a range of the records in the buffer in the format
specified by the selections in the Configuration.
At the end of the Buffer Report, the following summary information is
printed.
• Data displayed in Cumulative or Differential mode
• Raw (“Particle Counts”) or Normalized (“Particles/ft
3
” or
“Particles/m
3
”)
• # of samples in the report
• For each channel size:
• Maximum Value
• Minimum Value
• Average
• Standard Deviation
The UCL calculation will be printed for each channel size if the num-
ber of locations in the range is 2 through 9. The calculation will only
print if the Format settings (Setting screen) are set to CUML and
NORM. Any other format settings or locations less than 2 and greater
than 9 will print N/A for each channel.
View the same data in different formats just by changing the Particle
Display options on the Sample Setting screen.
The Data Summary can be viewed as:
• Raw Counts, Differential Data
• Raw Counts, Cumulative Data
• Normalized Counts, Differential Data, Particles/ft
3
• Normalized Counts, Differential Data, Particles/m
3
• Normalized Counts, Cumulative Data, Particles/ft
3
• Normalized Counts, Cumulative Data, Particles/m
3