Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual
4-46 5HY
PRINT RANGE (BUFFER REPORT)
The instrument stores up to 3000 records in its data buffer. The Buffer
Report prints all or a range of the records in the buffer in the format
specified by the selections in the Configuration.
At the end of the Buffer Report, the following summary information is
printed.
Data displayed in Cumulative or Differential mode
Raw (“Particle Counts”) or Normalized (“Particles/ft
3
” or
“Particles/m
3
”)
# of samples in the report
For each channel size:
Maximum Value
Minimum Value
Average
Standard Deviation
The UCL calculation will be printed for each channel size if the num-
ber of locations in the range is 2 through 9. The calculation will only
print if the Format settings (Setting screen) are set to CUML and
NORM. Any other format settings or locations less than 2 and greater
than 9 will print N/A for each channel.
View the same data in different formats just by changing the Particle
Display options on the Sample Setting screen.
The Data Summary can be viewed as:
Raw Counts, Differential Data
Raw Counts, Cumulative Data
Normalized Counts, Differential Data, Particles/ft
3
Normalized Counts, Differential Data, Particles/m
3
Normalized Counts, Cumulative Data, Particles/ft
3
Normalized Counts, Cumulative Data, Particles/m
3