9.22 Using Sequences
Using sequences from remote mode is very simple and the basic steps are outlined below.
Start by sending an MC to clear the results memory. It can be a good idea to download
the sequence to the LA101 before running it to be sure of what you are running, and it is
best downloaded as sequence 0 using a
1DS0,xxx... command to leave the manual mode
sequences (1-10) alone. The
SP command can be used to select the LPF or change the
timeout although most programs will probably leave the default values in use. Send SQ
to both units to run the sequence (or
SQ,m to run a single segment or SQ,,c to run on a
single channel etc). It is good practice to check the error state (using
E?) after running
the sequence to ensure that all of the segments were received correctly and that the unit
did not time out. Use SR? to read the sequence results, and then 2S? to retrieve the full
frequency response data if required (see section 9.16).