Level results are available. Since noise and crosstalk measure-
ments are relative to the measured test level these segments must
be preceded by a test level segment (eg the sequences RCIN or
RCINT will generate this error. They should read TRCIN).
TOO LONG (64) Remote command line too long (over 250 characters).
TOO MANY Too many tolerance strings have been entered into the tolerance
editor for a segment which does not produce that many results
values.
TOO MANY > CALLS Sequence may call other sequences up to 10 levels deep (see
section 4.23). Sequences should not call themselves recursively,
otherwise this error will occur (use << instead to repeat the entire
sequence).
TOO MANY TONES Only 32 tones (96 parameters) are allowed in a ! tone burst
segment. More can be accommodated by using two ! segments
one after the other (using different sequences if necessary − see
sections 4.16 and 4.23).
TOO QUICK (68) Missed start bit of remote data (baud rate may be incorrect, or
there may not have been enough time allowed to execute the
previous command).
UNUSED The measurement option is not defined and should not be used.
181
Appendix E. Error Messages