USE AND MAINTENANCE MANUAL DLC
NET
Series
Use as interface with indicator
LOGIC S.r.l. - M0145Db.docx
Mod. L0006A01 of 30/05/2016
Page 50 of 78
At this point the instrument will show the written and the
procedure is finished.
5.4.5 Sensitivity correction
( )
The procedure can be performed with the following steps:
Enter the menu, press the button until the menu
and press to enter the menu.
Press at any time to cancel and return to the previous menu.
The display shows the message , because the system waits for
input the value of data weight to correct.
Press to start editing the data. It is displayed the actual weight, with
the first digit flashing.
Use the button to increase the digit and to move on to
editing the next one. Press to confirm.
The display shows the message , the value that the instrument
should be displayed instead of the entered in .
Use the buttons and to increase the digit and move to the
next. Press to confirm.
At this point the instrument will show the written and the
procedure is finished.
5.4.6 Zero scale calibration
( )
The procedure can be performed with the following steps
From the main screen, press and move inside the top-level menu
with the button to the menu. Press to enter.
Press at any time to cancel and return to the previous menu.
The instrument will display the actual weight. Press again to edit
the value of zero.
At this point the display will show , with the first digit flashing,and
the instrument waits for the value of the weight added to the sample.
Enter the value of to be reset using the button to increase
the digit and to move on to editing the next one. Press to
confirm.