43
SD 320_2 05/2015
Menu Parameter Values Meaning
rEAd Logg: Review stored measurements, see Chapter 8.1: Manual
recording ("Func-Stor"). If data in the single value logger (Logger:
"Func Stor) is first shown as "rEAd Logg" menu: see also Chapter 8
Data logger for this purpose.
Set Configuration: General settings
Input: Selection of the measurement
**
Set Choice of displayed unit via set-key
Attention: Causes limitations for alarm
und analog output (DAC)
Cond Conductivity
rES: Specific resistance
SAL Salt content/salinity
TDS Total Dissolved Solids
TDS measurement: Conversion factor .
(only with Inp = TDS)
0.40 - 1.00 Conversion factor for TDS measurement
Cell Range: Adjustment of the cell constant: Cell con-
stant range
0.01 Purest water, electrodes with K ~ 0.01
0.1 Purest water, electrodes with K ~ 0.1
1
Standard electrode in the standard delivery
contents, e.g.
SET1 K = 0.55
SET2 K = 0.42
10 Electrode with K=10
Cell Factor: Adjustment of the cell constant:
Multiplication factor
0.3800 -
1.5000
Multiplication factor of the cell constant:
Cell constant CELL = CELL Range
.
CELL
Factor
t-Input: Selection of the temperature input
NTC NTC 10k sensor (LC 12)
Pt Pt1000 sensor (LC 16)
Range: Selection of the display range (conductivity,
spec. resistance or TDS)
Auto Automatic range selection
e.g.
0.0 ... 500.0
µS/cm
Example for CELL rAng 1 and InP Cond:
for others, see chap. 6.3 - 6.5
...
0 - 1000 mS/cm
Example for CELL rAng 1 and InP Cond:
for others, see chap. 6.3 - 6.5
(**) If logger is running parameters marked with (**) cannot be called.