Performing a Quality Control test
• Place the Instrument on a level, stable surface before starting a test.
• Use only LumiraDx Quality Controls.
• Apply Quality Control solution AFTER inserting the Test Strip and the
Instrument prompt.
• Do NOT apply more than one drop of Quality Control solution.
• Close the Instrument door after applying the Quality Control solution and
the Instrument prompt.
• Do NOT touch the Test Strip until the test has finished and the result
isdisplayed.
• Do NOT open the door or move the Instrument during the test or an error
will result and the test result will not be saved.
• Dispose of all Test Strips used for Quality Control testing safely in
accordance with local regulations and procedures.
• Follow the test operating conditions in the LumiraDx Test Strip Product
Insert.
• Follow the information on correct handling of Test Strips in the LumiraDx
Test Strip Product Insert.
Cleaning and disinfecting
• Always wear gloves whilst cleaning and disinfecting the Instrument.
• Always use LumiraDx recommended materials to disinfect the Instrument.
Details of LumiraDx recommended materials can be found at lumiradx.com.
• CAUTION: Excessive liquid may damage the Instrument. It is important
for the protection of the Instrument that exposure to excess moisture is
prevented. All disinfection cloths and/or wipes should only be slightly damp,
with any excess liquid being manually removed from the cloth or wipe
before use.
• CAUTION: Liquid should not be placed in or near the vents on the bottom
or back of the Instrument.
• CAUTION: Alcohol wipes alone are not sufficient to disinfect the
Instrument for blood-based samples, due to the potential presence
ofbloodborne pathogens.
• Do NOT spray or pour solution directly onto the Instrument.
• Following cleaning and disinfection, and before performing a patient
test, change gloves and wash hands.