Instrument Operation - 55
4.9 Performing a Quality Control Test
This Platform User Manual describes the basic Quality Control test method for the
LumiraDx platform. If there are differences in the test method for a particular QC test, the
Instrument will prompt at each step of the process. Please refer to the LumiraDx Test Strip
Product Insert before performing a new QC test for specific information.
Important Notes
ALWAYS
Wear a clean pair of gloves before performing a QC test.
Place the Instrument on a level, stable surface before performing a test.
Use only LumiraDx Quality Controls.
Apply Quality Control AFTER inserting the Test Strip and the Instrument
prompt.
Follow the test operating conditions in the LumiraDx Test Strip Product
Insert.
Follow the information on correct handling of Test Strips in the LumiraDx
Test Strip Product Insert.
To view test information during a QC test, or on the test result screen, tap the at the
top of the touch-screen.
To cancel a QC test tap ‘Cancel’ at the top of the touch-screen and follow the
Instrument prompts.