Instrument Operation - 47
7. Open Door and Insert Test Strip
The Instrument display will prompt to open
the door and insert a Test Strip.
Remove a Test Strip from its packaging and
hold with the blue label side facing upward.
Line up the black alignment rib on the Test
Strip with the black line on the Instrument.
Insert the Test Strip gently as far as it will go.
The Instrument will sound when the Test Strip
is detected.
Do NOT touch Test Strip Sample
Application Area.
Do NOT touch Test Strip contacts.
Do NOT apply sample until prompted.
8. Select Sample Type
Select the intended sample type from the list
displayed.
Sample type is test type dependent (refer to
the LumiraDx Test Strip Product Insert for more
information).
Note: The Instrument will prompt to install the
Lot Calibration File when inserting a new Test
Strip Lot.
Note: Capillary Blood refers to both finger
stick and transfer tube applications.