Cleaning and Disinfecting - 71
Cleaning Instrument touch-screen Cleaning the area around the Test Strip slot
Cleaning and disinfecting
Always wear gloves whilst cleaning and disinfecting the Instrument.
Always use LumiraDx recommended materials to disinfect the
Instrument. Details of LumiraDx recommended materials can be found
at lumiradx.com.
CAUTION: Excessive liquid may damage the Instrument. It is
important for the protection of the Instrument that exposure to excess
moisture is prevented. All disinfection cloths and/or wipes should only
be slightly damp, with any excess liquid being manually removed from
the cloth or wipe before use.
CAUTION: Liquid should not be placed in or near the vents on the
bottom or back of the Instrument.
CAUTION: Alcohol wipes alone are not sufficient to disinfect the
Instrument for blood-based samples, due to the potential presence of
bloodborne pathogens.
Do NOT spray or pour solution directly onto the Instrument.
Following cleaning and disinfection, and before performing a patient
test, change gloves and wash hands.