5.2.3 Disinfection guidance – if only testing swab-based samples
It is recommended to disinfect the Instrument at least once per day, or if contamination
is suspected.
CAUTION: Excessive liquid may damage the Instrument. It is important for
the protection of the Instrument that exposure to excess moisture is prevented.
Alldisinfection cloths and/or wipes should only be slightly damp, with any excess liquid
being manually removed from the cloth before use.
Procedure:
1) Using a LumiraDx recommended disinfecting material, wipe the external surfaces
of the Instrument while taking care to avoid the door hinges, Test Strip inlet, back and
bottom vents, power cord, and USB port.
2) Allow the disinfectant at least 1 minute contact time before testing the next sample.
3) Dispose of disinfectant materials in accordance with local biohazardous waste
disposal procedures.
The Instrument is now ready to perform another test.
For technical assistance or questions and information about LumiraDx approved
materials go to lumiradx.com.