r-net user guide
36 5940A R-net User Guide DCR1266
magicmobility.com.au
Chapter 3 - Diagnostics
1 INTRODUCTION
The primary objective of this chapter is to assist service personnel in finding the likely area of a detected trip
within the whole wheelchair electrical system. It is important to realize that even though the control system is
signalling a trip, it may not be the control system itself that is defective. This is because the control system is
able to detect problems in other electrical components (motors, batteries, solenoid brakes, etc.) or, more
importantly, the wiring to them. When a control system has detected a trip, a system trip is indicated.
This chapter covers the diagnostic process for an LCD based JSM. If a JSM-LED is being used, refer to section
Control System Status Indication in Chapter 2.
WARNING:
Diagnostics should only be conducted by healthcare professionals with in-depth knowledge of R-Net
electronic control systems. An incorrect or badly effected repair could result in an unsafe set-up of a
wheelchair.
1.1 Diagnostic Procedure
To diagnose a trip please follow this procedure:
Read and note the Trip Text displayed, the Identified Module and the Trip Code. Refer to section 2.
Switch off the control system.
Make sure that all connectors on the listed Module and the wheelchair are mated securely.
Check the condition of the battery.
Find the definition of the Trip Text, and take the required action. Refer to section 3.
Switch on the control system again and try to drive the wheelchair. If the safety circuits operate again,
switch off and do not try to use the wheelchair.
Contact your service agent.
2 DIAGNOSTIC SCREENS
2.1 Current diagnostic screen
When the control system safety circuits have operated and the control system has been prevented from
moving the wheelchair a diagnostics screen will be displayed.
This indicates a system trip, i.e. the R-net has detected a problem somewhere in the wheelchair’s electrical
system.
If the error is in a non-active module, for example in the ISM but with a Drive Mode selected, then drive will
still be possible, however, the diagnostic screen will appear intermittently.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Magic Mobility R-NET CJSM2 and is the answer not in the manual?

Magic Mobility R-NET CJSM2 Specifications

R-Net 80

Power IconPower
supply voltage24 Vdc
operating voltage16 Vdc to 35 Vdc
peak voltage35 Vdc
Current IconCurrent
brake current200 µA min. 1 A max.
battery charging current12 Arms max.
maximum drive current80 A
Environmental Specifications IconEnvironmental Specifications
moisture resistanceIPX4
operating temperature-25°C to +50°C
storage temperature-40°C to +65°C

R-Net 90

Power IconPower
supply voltage24 Vdc
operating voltage16 Vdc to 35 Vdc
peak voltage35 Vdc
Current IconCurrent
brake current200 µA min. 1 A max.
battery charging current12 Arms max.
maximum drive current90 A
Environmental Specifications IconEnvironmental Specifications
moisture resistanceIPX4
operating temperature-25°C to +50°C
storage temperature-40°C to +65°C

R-Net 120

Power IconPower
supply voltage24 Vdc
operating voltage16 Vdc to 35 Vdc
peak voltage35 Vdc
Current IconCurrent
brake current200 µA min. 1 A max.
battery charging current12 Arms max.
maximum drive current120 A
Environmental Specifications IconEnvironmental Specifications
moisture resistanceIPX4
operating temperature-25°C to +50°C
storage temperature-40°C to +65°C