r-net user guide
41 5940A R-net User Guide DCR1266
magicmobility.com.au
3.17 PM Memory error
This is a specific Power Module based trip.
Check all cables and connections.
Using the R-net PC Programmer, re-program the control system.
This should be done with either the most current specific program file for the wheelchair or the
manufacturer’s original programming file.
WARNING:
Programming should only be conducted by healthcare professionals with in-depth knowledge of R-Net
control systems. Incorrect programming could result in an unsafe set-up of a wheelchair for a user.
If the trip is still present after the above checks have been made, then the PM may be defective. Refer to
Section 5.
3.18 Bad cable
This occurs when the control system detects a fault in the wiring in the communication cables between any of
the modules.
Check all cables and connections for continuity.
If there is any visible damage to cables, replace and cycle power.
Disconnect one cable from the system at a time cycling the power after each disconnection.
If the trip is still present after the above checks have been made, then the PM may be defective. Refer to
Section 5.
3.19 Bad settings
This occurs when the control system detects incorrect or invalid program settings.
Check all parameter settings and re-program the control system using the R-net PC Programmer.
Make a note of the current parameter settings and then reset the control system to default settings.
Re-program the required settings in small groups, cycling the power after each group to see if the trip
occurs.
If the trip is still present after the above checks have been made, then the PM may be defective. Refer to
Section 5.
WARNING:
Programming should only be conducted by healthcare professionals with in-depth knowledge of R-Net
control systems. Incorrect programming could result in an unsafe set-up of a wheelchair for a user.
3.20 Module error
This occurs when the control system detects a trip within a specific module. The module will be identified on
the diagnostics screen.
Check all cables and connections.
Cycle the power.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Magic Mobility R-NET CJSM2 and is the answer not in the manual?

Magic Mobility R-NET CJSM2 Specifications

R-Net 80

Power IconPower
supply voltage24 Vdc
operating voltage16 Vdc to 35 Vdc
peak voltage35 Vdc
Current IconCurrent
brake current200 µA min. 1 A max.
battery charging current12 Arms max.
maximum drive current80 A
Environmental Specifications IconEnvironmental Specifications
moisture resistanceIPX4
operating temperature-25°C to +50°C
storage temperature-40°C to +65°C

R-Net 90

Power IconPower
supply voltage24 Vdc
operating voltage16 Vdc to 35 Vdc
peak voltage35 Vdc
Current IconCurrent
brake current200 µA min. 1 A max.
battery charging current12 Arms max.
maximum drive current90 A
Environmental Specifications IconEnvironmental Specifications
moisture resistanceIPX4
operating temperature-25°C to +50°C
storage temperature-40°C to +65°C

R-Net 120

Power IconPower
supply voltage24 Vdc
operating voltage16 Vdc to 35 Vdc
peak voltage35 Vdc
Current IconCurrent
brake current200 µA min. 1 A max.
battery charging current12 Arms max.
maximum drive current120 A
Environmental Specifications IconEnvironmental Specifications
moisture resistanceIPX4
operating temperature-25°C to +50°C
storage temperature-40°C to +65°C