MAGNAMED TECNOLOGIA MÉDICA S/A
Specifications of Maintenance and Calibration
Table 28 - Maintenance and calibration specifications
Description
Specifications
Tolerance
Review and replacement of
exhalation valve’s diaphragm
Under inspection or 5,000 hours or 12 months
(whichever occurs first)
500 h / 1
months
Review and replacement of the
galvanic cell O
2
Recommended replacement if there are problems
in the calibration or 10,000 hours or 24 months
(whichever occurs first)
Review and replacement of the
internal batteries
10.000 hours or 24 months (whichever occurs first)
Equipment Review
5.000 hours or 12 months (whichever occurs first)
Equipment Calibration
5.000 hours or 12 months (whichever occurs first)
Expiration date
INDEFINITE
(1)
---
(1) The expiration date is undetermined when respected and implemented regular reviews
Electromagnetic Compatibility
Changes or modifications to this equipment not expressly approved the MAGNAMED can cause EMC
issues with this or other equipment. Contact the MAGNAMED to receive technical assistance.
This equipment has been designed and tested to comply with the EMC standards applicable as
described below:
Immunity: IEC 60601-1-2
Emission: CISPR11
Approvals: OS/IEC 60601-1
The use of cell phones or other transmitting equipment of radio frequency (RF) near the system
may cause unexpected or adverse outcomes. Monitor the operation if there is radio emission
sources in the vicinity
The use of other electrical equipment on or around the system may cause interference. Before
use in the patient, you should check that the equipment usually works in the defined
configuration.