MAGNAMED TECNOLOGIA MÉDICA S/A
Specifications of performance
Table 26 - Performance Specifications
Parameter
Specifications
Unit
Tolerance
Valves Response Time T
0.90
5
ms
20%
Maximum Flow in Pressure Support
and Spontaneous Breathing
180
L/min
10%
Maximum flow leakage compensated
Neonate
20
L/min
10%
Pediatric and
Adult
50
L/min
10%
Controlled ventilation pressure is recommended for larger leak flows than the limit specified above.
In this case the maximum flow offset can be greater than 100 L / min.
Specifications of Expiratory Limb Resistance
Table 27 - Branch resistance specifications expiratory
Breathing
circuit
Flow
(L/min)
Expiratory resistance (hPa or cmH
2
O)
Circuit
Circuit + Flow
Sensor
Circuit + Flow
Sensor + Filter
HME
Circuit + Flow
Sensor + CO
2
Sensor+ HME
Filter
Neonate
5,0
0,6
1,7
Pediatric
30,0
0,4
3,4
4,1
4,3
Adult
60,0
0,8
1,4
3,1
3,5
Adult
60,0
3,8
4,4
6,1
6,5