37
MAGNAMED TECNOLOGIA MÉDICA S/A
6.5.3 CONFIG
CONFIG menu is available in all models.
6.5.3.1 Auxiliary resources in ventilation
Nebulizer
Tracheal gas insufflation (TGI)
Non invasive ventilation (NIV)
The nebulizer flow is synchronized with the inspiration (inspiratory flow) and has resources of volume
compensation and FiO2.
The TGI flow is synchronized with expiration and activated when the exhaled flow reaches 25% of the
maximum peak flow expired. Its end is determined by the beginning of the next inspiratory cycle.
6.5.3.2 Pressure measurement units
The operator may choose to display the values of pressure in mbar, hPa or cmH2O, as follow:
1 mbar (milibar) = 1 hPa (hectoPascal) = 1.016 cmH
2
O (centimeter of water)
In a practical way, these units are equivalent, and it will be adopted:
1 mbar = 1 hPa ≈ 1 cmH
2
O
To change the unit of pressure, just tap on the corresponding button on the SETUP screen.
It is also possible to adjust the high pressure units used for measuring the pressure of gases
in the equipment inlet.
6.5.3.3 Language
The ventilator interface may be set up to three different languages. To change the language,
simply tap on the corresponding button on the SETUP screen.
6.5.3.4 Patient change
The patient selection shall set a previous configuration of ventilation according to the following
table: