MAGNAMED TECNOLOGIA MÉDICA S/A
Monitored Parameter Specifications
Table 19- Monitored Ventilatory Parameters
Parameter
Range
Resolution
Tolerance
(1)
Instant Pressure
0 to 120 cmH
2
O
1
1cmH
2
O or
2% of read
Peak Pressure
0 to 120 cmH
2
O
1
1cmH
2
O or
2% of read
Mean Pressure
0 to 120 cmH
2
O
1
1cmH
2
O or
2% of read
Plateau Pressure
0 to 120 cmH
2
O
1
1cmH
2
O or
2% of read
PEEP
0 to 120 cmH
2
O
1
1cmH
2
O or
2% of read
Intrinsic PEEP (iPEEP)
0 to 120 cmH
2
O
1
1cmH
2
O or
2% of read
Measured Flow
-5,0 to 5,0 L/min
0,1
50mL/min or
2% of read
-20,0 to -5,0 and 5,0 to 20,0
L/min
0,2
-180 to -20 and 20 to 180
L/min
1
Tidal Volume
(2)
0 to 999 mL
1
2,5mL or
5% of the measured
value
1,00 to 3,00 L
0,01
Minute Volume (MV)
0,001 to 0,999L
0,001
0,18L or
3% of the measured
value
1,0 to 9,99 L
1,0
10,0 to 9,99 L
10,0
Inspiratory Time
0,05 to 9,99 s
0,01
0,01s
10,0 to 60,0 s
0,1
0,1s
Expiratory Time
0,05 to 9,99 s
0,01
0,01s
10,0 to 60,0 s
0,1
0,1s
Ratio I:E
1:599 a 599:1
1:0,1
2%
Total Breathing Rate
0 to 200 bpm
1
1bpm or
1% of the measured
value