11.17.4 Electrical safety
The following are the precautions that should be observed when combining these items (not medical
equipment) with the system.
Items that do not comply the requirements of IEC 60601-1 should not be placed within 1.5 m of
the patient.
All items (electromedical equipment or not) connected to the system output signal cable / input
must receive power from a source of alternating current using separate transformer (according
to IEC 60989) or have protective conductor additional land.
Do not connect directly electrical equipment nonmedical the AC wall outlet. Using AC power
supply with own transformer. Otherwise, the leakage current increase above the levels accepted
by the IEC 60601-1 under normal terms and conditions of a single failure. This may cause
dangerous electrical shock to the patient or operator.
After connecting any equipment into these outlets, subject the system to a complete test leakage
current (according to IEC 60601-1).
The electromedical system operator shall not touch electrical equipment no doctor and the
patient simultaneously. This may cause dangerous electrical shock to the patient or operator.