EasyManua.ls Logo

Magnamed Fleximag - Adjustable Parameter Specifications

Magnamed Fleximag
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
87
MAGNAMED TECNOLOGIA MÉDICA S/A
Adjustable Parameter Specifications
Table 18 Adjustable Parameters
Parameter
Specification
Resolution
Unit
Tidal Volume
10 to 3000
Pediatric
10 to 100: 5
mL
100 to 300: 10
Adult
100 to 1000: 10
1000 to 3000: 50
Breathing Rate
(1)(2)
0 to 200
Neonatal
0 to 200:1
rpm
Pediatric
0 to 200:1
Adult
0 to 100:1
Rise Time
0 to 2,0
0.1
s
Pause
0 to 70
10
%
0 to 2,0
0,1
s
Inspiratory and Limit Pressure
0 to 120
1
cmH
2
O
∆PS
0 to 120
1 cmH
2
O
cmH
2
O
PEEP
0 to 50
1 cmH
2
O
cmH
2
O
Pressure trigger
0,0 to -20
0.00 to -2.0 : - 0,2
cmH
2
O
-0.05 to -20 : - 1
Flow trigger
0,0 to 30
0.00 to 1.0 : 0,1
L/min
1,0 to 30,0 : 0,5
Cycled-Flow
5 to 80
(max. 3 s)
5
%
O
2
Concentration
21 to 100
1
% vol
Inspiratory Time
0,05 to 30
0.05 to 0,70 : 0,01
s
0.70 to 1,00 : 0,05
1,0 to 30,0 : 0,1
Inspiratory Flow’s Waveform
Square,
Descending or
Decelerated,
Ascending or
Accelerated,
Sinusoidal or sinusoid
---
---
CPAP
0 to 50
1
cmH
2
O
High Pressure (APRV/DualPAP)
5 to 90
1
cmH
2
O
Low Pressure (APRV/DualPAP)
0 to 45
1
cmH
2
O
High Time (APRV/DualPAP)
0,10 to 59,9
0,10 to 0,70: 0,01
s
0,70 to 1,00: 0,05

Table of Contents

Related product manuals