Operation Manual
Mechanical Ventilator
Fleximag (display 15”)

Table of Contents

Questions and Answers

  • S
    Sheri VillarrealAug 21, 2025
    What to do if the disconnection alarm is on for my Magnamed Fan?
    • K
      kristy44Aug 22, 2025
      The disconnection alarm on your Magnamed Fan can be triggered by several issues. First, check the respiratory circuit for any disconnections and ensure all connections are firm. If there's a lack of inspiratory flow, verify the flow and increase it if needed. An alteration of the patient’s respiratory mechanics might also be the cause, requiring you to establish new parameters for ventilatory support. Inspect the exhalation valve diaphragm for incorrect mounting or damage; reposition it correctly or substitute the diaphragm if necessary. If none of these resolves the issue, it may indicate a failure in the pressure control electronic system, and you should request technical assistance.
  • J
    jonathanhaleAug 25, 2025
    How to fix a low battery alarm on a Magnamed Fan?
    • J
      James SaundersAug 25, 2025
      If you're getting a low battery alarm on your Magnamed Fan, it's likely due to the internal battery running low after being used without a power source. Immediately re-establish the equipment's connection to the power supply, or turn off the equipment and provide another means of ventilatory support to the patient.
  • M
    Mary PettyAug 26, 2025
    Why is the power source alarm active on my Magnamed Fleximag?
    • C
      Corey WilliamsAug 26, 2025
      The power source alarm on the Magnamed Fan indicates either a disconnection from the electric power cord or a failure in the power grid. Immediately re-establish the equipment connection to the power supply or use the equipment with the internal battery for transport. If the issue is a failure in the power grid, re-establish the power supply.
  • D
    Derek BrooksAug 28, 2025
    What causes the low pressure alarm on a Magnamed Fan?
    • G
      Gabrielle SparksAug 28, 2025
      The low pressure alarm on your Magnamed Fan can be triggered by a couple of things. It could be due to an alteration of the patient’s respiratory mechanics, requiring you to establish new parameters for ventilatory support. Another possibility is excessive leakage in the respiratory circuit, so locate the leakage and correct it.
  • C
    Calvin JacksonAug 29, 2025
    Why is the audible alarm inoperative on my Magnamed Fan?
    • G
      George WilsonAug 29, 2025
      The audible alarm on your Magnamed Fan might not be working because the mute alarm is active. Turn off the mute alarm. If that's not the case, it could be an electronic failure, and you should request technical assistance.
  • S
    Sara FloresSep 1, 2025
    What does the communication failure alarm mean on my Magnamed Fleximag Fan?
    • M
      Mr. Leslie LoganSep 1, 2025
      The communication failure alarm on the Magnamed Fan indicates an electronic failure, and you should request technical assistance.
  • N
    Nancy MitchellSep 2, 2025
    How to start ventilation on Magnamed Fleximag?
    • S
      Sylvia WhiteSep 3, 2025
      If your Magnamed Fan doesn't start ventilation, it might be in STAND BY mode. Press the STAND BY button for one second to start the ventilation.
  • S
    susanburchSep 5, 2025
    What causes incorrect PEEP control on a Magnamed Fleximag Fan?
    • W
      william04Sep 5, 2025
      If you're experiencing incorrect PEEP control with your Magnamed Fan, it could be due to exhalation valve calibration issues. Try restarting the equipment and calibrating the exhalation valve.
  • X
    xkirkSep 7, 2025
    What to do if the exhalation valve test fails on my Magnamed Fan?
    • A
      Alice MartinSep 7, 2025
      If the exhalation valve test fails on your Magnamed Fan, it could be because the output of the breathing circuit was not occluded. Restart the ventilator and redo the self-test with the respiratory circuit closed. Another reason could be that the diaphragm of the exhalation valve is misplaced. Reposition the exhalation valve diaphragm, restart the machine, and redo the self-test.
  • D
    Donna PearsonSep 9, 2025
    Why can't I activate the automatic adjust of the alarms on my Magnamed Fan?
    • G
      Gail WashingtonSep 9, 2025
      If you cannot activate the automatic adjustment of the alarms on your Magnamed Fan, it is probably because the equipment is in STAND BY mode. The automatic adjustment of the alarms may only be calculated if the ventilator is cycling. Press the STAND BY button for 1 second to start the ventilation and wait for the ventilation stabilization to activate the automatic adjustment.

Summary

Safety Annotations

Definitions

Explains key terms and concepts within the safety section.

Warning

Alerts users to potential hazards, serious adverse reactions, and equipment risks.

Attention

Provides important information regarding equipment use, maintenance, and potential issues.

Observation

Offers important supplementary information and notes for user guidance.

Characteristics

Intended Use

Defines the ventilator's purpose and intended patient population for its use.

Functioning Principle

Explains the operational basis and modular components of the FlexiMag ventilator.

General Characteristics

Details the overall features, capabilities, and specifications of the FlexiMag ventilator.

Safety Characteristics

Outlines the built-in safety mechanisms, protective features, and alarm systems of the ventilator.

Unpacking the Product

Initial Verifications

Guides users through essential checks to ensure the equipment's integrity after delivery.

Relation of Components of FlexiMag 15”

Lists and describes the integral parts and accessories included with the FlexiMag ventilator.

Component Identification

Front View

Identifies and labels key components and controls visible on the front panel of the ventilator.

Back View

Identifies and labels key components and connections located on the rear panel of the ventilator.

Preparation for Use

Assembly

Provides step-by-step instructions for the correct physical assembly of the ventilator unit.

Connection to Power Supply

Explains the procedure for safely and correctly connecting the ventilator to an electrical power source.

Verifications Before Use

Details the mandatory pre-operational checks and tests required before patient use.

Use Instructions

Initial Sequence

Outlines the initial steps for powering on and preparing the ventilator for patient connection.

Adjustment Buttons

Describes how to operate the adjustment buttons for setting ventilation parameters and alarms.

Main Screen

Explains the layout and main functional areas of the ventilator's primary user interface screen.

Magnamed Fleximag Specifications

General IconGeneral
BrandMagnamed
ModelFleximag
CategoryFan
LanguageEnglish

Related product manuals