85
MAGNAMED TECNOLOGIA MÉDICA S/A
11.10.9 APRV
APRV – Airway pressure release ventilation
Description:
This mode allows spontaneous cycles on 2
levels of baseline pressure and can be
achieved by appropriate adjustments in
DualPAP mode;
For this mode selects inverted ratio in
DUALPAP. With this adjustment is carried out
a pressure relief airway, obtaining APRV –
Airway Pressure Release Ventilation
Set parameters:
• PR. SUPERIOR
• T. SUPERIOR
• PR. INFERIOR
• T. INFERIOR
• CONCENTRATION
• ΔPS (Pressure Support – PEEP)
• PRESSURE LIMIT;
• CYCLING FLOW (% FLOW);
• TRIGGER PRESSURE
• FLOW PRESSURE
• RISE TIME
• FLOW (NEONATAL)
• BACKUP
o Backup VCV
▪ VOLUME
▪ RATE
▪ FLOW
▪ LIMIT PRESSURE
o Backup PCV
▪ PRESSURE INSP
▪ RATE
▪ TIME INSP
▪ RISE TIME
o Backup PLV-NEONATAL
▪ PRESSURE INSP
▪ RATE
▪ TIME INSP
o Backup Auto
(1)
1 – Whenever the apnea set
time is reached, the ventilator
triggers a ventilatory cycle
whose configuration is based
on the current mode settings.