MAGNAMED TECNOLOGIA MÉDICA S/A
11.10.3 PLV
PLV Pressure Limited Ventilation
Description:
In this ventilation mode of continuous flow,
secure the respiratory rate, inspiratory time
and inspiratory pressure limit.
The trigger, if the sensitivity setting is
disabled, is determined exclusively according
to the respiratory rate, but the cycling
happens according to the inspiratory time.
The current volume becomes dependent on
the preset inspiratory pressure, the
impedance conditions of the respiratory
system and inspiratory time selected by the
operator.
Normally to observe the flow curve, we see a
peak flow will decrease as the time passes.
Set parameters:
PRESSURE INSPIRATORY
FREQUENCY
TIME INSPIRATORY
PEEP
CONCENTRATION
FLOW ( )
TRIGGER PRESSURE
TRIGGER FLOW
Note: automatic Backup
(1)
1 Whenever the apnea set time is reached, the ventilator
triggers a ventilatory cycle whose configuration is based on
the current mode settings.