MAGNAMED TECNOLOGIA MÉDICA S/A
their use.
Safety Characteristics
• Anti-asphyxia valve to protect against failures in gas supply.
• Relief valve of 100 hPa, in compliance with basic standard of ventilators, avoiding possible overpressure in the respiratory
circuit.
• Active overpressure valve to detect obstructions is activated to reduce the pressure in the patient circuit.
• The equipment has audiovisual alarms related to both equipment operating condition (technical alarms), as the ventilation
process.
• Backup energy system that allows the equipment to run using batteries for about 3,5 hours
.
• Speaker for alarms and alerts.
• High brightness RED LED for prompt alarm identification.
• Possibility of automatic adjustment of alarms.
• Option of autodiagnosys to check alarms, failures detection, leakage measurement, system resistance and compliance.
• Use of the equipment even in case of single gas failure (O2 or compressed air)