MAGNAMED TECNOLOGIA MÉDICA S/A
11.10.5 V-SIMV
V-SIMV Volume Synchronized Intermittent Mandatory Ventilation
Description:
In V-SIMV, secure respiratory rate, tidal
volume and the inspiratory flow or the ratio
or inspiratory time, beyond the sensitivity
criteria for the occurrence of ventilator
trigger by the patient.
This mode allows the ventilator to apply the
mandatory cycles predetermined in sync
with the inspiratory effort of the patient.
The mandatory cycles occur in the
predetermined time window (according to
the set respiratory rate), but synchronized
with the patient trigger.
If apnea, the next cycle will be triggered by
time until they return the inspiratory
incursions patient
To obtain this IMV mode, just off the
support pressure support setting the zero
pressure (ΔPS = 0) or the flow and pressure
sensitivities equal to zero.
Set parameters:
VOLUME
RATE
FLOW or RATIO or TIME INS
PEEP
CONCENTRATION
FLOW WAVEFORM
PAUSE (%)
TRIGGER PRESSURE
TRIGGER FLOW
ΔPS (Pressure Support - PEEP)
RISE TIME
FLOW CYCLING (% FLOW)
LIMIT PRESSURE
Note: automatic Backup
(1)
1 Whenever the apnea set time is reached, the ventilator
triggers a ventilatory cycle whose configuration is based on
the current mode settings.