47
MAGNAMED TECNOLOGIA MÉDICA S/A
The distal flow sensor must not be sterilized by steam sterilization.
For the internal disinfection of DISTAL flow sensor, do not use tools that may generate
mechanical forces, such as air or water jet, at the risk of damaging the filament.
Drying of DISTAL flow sensor should occur naturally in ambient air, so avoid the use of
compressed air or dryer.
If there is a possibility of the sensor remain infected, replace it promptly.
Sterilization
The components that get in touch with the respiratory gases must be removed for cleaning and
sterilization;
Do not use abrasive agents to carry out cleaning;
Do not use alcohol to clean the plastic parts;
Do not immerse the equipment in any liquid.
8.3.1 Steam sterilization
Perform the set sterilization as the autoclave manufacturer's recommendation.
Table 8 Autoclavable accessories
Description
Autoclave cycles
Breathing circuit
50
Silicon line
50
Exhalation valve
50
Diaphragm of exhalation valve
50
Important Advices