Operation Manual
Mechanical Ventilator
Fleximag (display 15”)

Table of Contents

Questions and Answers

  • E
    Elizabeth ChavezSep 12, 2025
    What causes the test failure of distal flow sensor on a Magnamed Fan?
    • K
      Karen WellsSep 12, 2025
      If the distal flow sensor test fails on your Magnamed Fan, it could be because the output of the breathing circuit was not occluded. Restart the equipment and redo the self-test with the respiratory circuit closed.
  • R
    Richard LarsonSep 12, 2025
    What does test failure of proportional valves mean on Magnamed Fan?
    • K
      Kelsey MontgomerySep 12, 2025
      If the proportional valves test fails on your Magnamed Fan, it's likely due to the pressure of the gas network being below the lower limit. Check and reset the network pressure to reach the specified range.
  • C
    christopher27Sep 15, 2025
    Why does the parameter setting return to the previous value on my Magnamed Fleximag?
    • B
      Brandon OsborneSep 15, 2025
      If the parameter setting returns to the previous value on your Magnamed Fan, it means that the adjustment was not confirmed. Confirm the adjustment by pressing the button or by touching the parameter.
  • J
    Julie BoydSep 17, 2025
    Why is the backup not active on my Magnamed Fleximag?
    • K
      Kendra SalazarSep 17, 2025
      If the backup is not active on your Magnamed Fan, the apnea time alarm might be off. Set an interval for the apnea alarm.
  • M
    michael46Nov 19, 2025
    Why is the high pressure alarm going off on my Magnamed Fan?
    • S
      Summer JonesNov 19, 2025
      The high pressure alarm on your Magnamed Fan may be triggered by several factors. It could be due to a change in the patient's respiratory mechanics, requiring you to establish suitable parameters for ventilatory support. Alternatively, there might be an obstruction in the respiratory circuit's expiratory limb or its exhalation valve; in this case, unblock the circuit or reposition the respiratory valve diaphragm. Another cause could be an obstruction in the patient’s airways, which would require you to remove obstructions or aspirate the patient’s airways. Check if the inspiratory pressure monitored is greater than expected, and adjust the setting of inspiratory pressure. Also, consider if the Automatic Tube Compensation (ATC) is on, and turn it off or set it to automatic compensation of ...

Summary

Safety Annotations

Definitions

Explains key terms and concepts within the safety section.

Warning

Alerts users to potential hazards, serious adverse reactions, and equipment risks.

Attention

Provides important information regarding equipment use, maintenance, and potential issues.

Observation

Offers important supplementary information and notes for user guidance.

Characteristics

Intended Use

Defines the ventilator's purpose and intended patient population for its use.

Functioning Principle

Explains the operational basis and modular components of the FlexiMag ventilator.

General Characteristics

Details the overall features, capabilities, and specifications of the FlexiMag ventilator.

Safety Characteristics

Outlines the built-in safety mechanisms, protective features, and alarm systems of the ventilator.

Unpacking the Product

Initial Verifications

Guides users through essential checks to ensure the equipment's integrity after delivery.

Relation of Components of FlexiMag 15”

Lists and describes the integral parts and accessories included with the FlexiMag ventilator.

Component Identification

Front View

Identifies and labels key components and controls visible on the front panel of the ventilator.

Back View

Identifies and labels key components and connections located on the rear panel of the ventilator.

Preparation for Use

Assembly

Provides step-by-step instructions for the correct physical assembly of the ventilator unit.

Connection to Power Supply

Explains the procedure for safely and correctly connecting the ventilator to an electrical power source.

Verifications Before Use

Details the mandatory pre-operational checks and tests required before patient use.

Use Instructions

Initial Sequence

Outlines the initial steps for powering on and preparing the ventilator for patient connection.

Adjustment Buttons

Describes how to operate the adjustment buttons for setting ventilation parameters and alarms.

Main Screen

Explains the layout and main functional areas of the ventilator's primary user interface screen.

Magnamed Fleximag Specifications

General IconGeneral
BrandMagnamed
ModelFleximag
CategoryFan
LanguageEnglish