EasyManua.ls Logo

Magnamed Fleximag - Use Instructions; Initial Sequence

Magnamed Fleximag
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26
MAGNAMED TECNOLOGIA MÉDICA S/A
Perform all verification procedures before each use
If any problem were identified, correct it BEFORE USING THE EQUIPMENT
If it was not possible to immediately correct the problem, call for authorized technical
assistance.
8
Check if gas inlet pressure is according to specification
Pressures above upper limit specification can damage the equipment
For inlet pressure below 250 kPa, the maximum flow will be 120 L/min
9
Verify a firm connection of the electric cord, when applicable. The ventilator can be used in
battery operation up to 210 minutes continuously under normal ventilation of the patient.
If the ventilator usage is extended in battery, an alarm occurs whose message
is LOW BATTERY, provide IMMEDIATE connection of the power supply to the
network power supply. If it was not possible, DISCONNECT the equipment
from the patient and provide appropriate means of ventilatory support.
10
If all items are check marked as OK, the equipment is ready for use.

Table of Contents

Related product manuals