MAGNAMED TECNOLOGIA MÉDICA S/A
Parameter
Specification
Resolution
Unit
1,00 to 59,90: 0,10
Low Time (APRV/DualPAP)
0,20 to 59,9
0,20 to 0,70: 0,01
s
0,70 to 1,00: 0,05
1,00 to 59,90: 0,10
I:E Ratio
1:599 to 299:1
(2)
1:0,1
---
Backup
(3)
OFF;PLV; PCV; VCV
and PRVC
---
---
Inspiratory Flow
1 to 180
1
L/min
Height
10 (neonatal) to 250
(adult)
(5)
1
cm
Nebulizer flow - 100% Oxygen
(4)
5 to 8 (not adjustable
directly)
---
L/min
Nebulizer time
1 to 50
1
min
TGI Flow 100% Oxygen
(4)
5 to 8 (not adjustable
directly)
---
L/min
Flow (PLV)
1 to 40
1
L/min
(1) Respiratory Rate 0 (zero) will only be achieved in spontaneous ways, with sensitivities and apnea time
alarm off.
(2) The minimum and maximum values of rate and I: E ratio, depend on the set ventilation mode.
(3) Adjustable backup options for spontaneous modes to other modes, backup is automatic.
(4) The nebulizer flow and TGI may not be activated simultaneously.
(5) The weight of the patient considered by the equipment is the ideal body weight, calculated according the
height of the patient.
FlexiMag ventilator attends any patients from preterm to morbidly obese, however, the patient
height adjustment used to calculate the ideal weight is limited.