MAGNAMED TECNOLOGIA MÉDICA S/A
Description:
In this mode, secure the respiratory rate,
tidal volume and inspiratory flow (or the
ration or inspiratory time).
The beginning of inspiration (trigger)
occurs according to the preset
respiratory rate. The shooting takes
place exclusively by time, if the
sensitivity setting is disabled.
The transition between inspiration and
expiration (cycling) occurs after the
release of the preset tidal volume at a
certain speed through the flow (or ratio
or inspiratory time).
Set Parameters:
VOLUME
FREQUENCY
FLOW or RATIO or TIME INS
PEEP
CONCENTRATION
LIMIT PRESSURE
PAUSE EXPIRATORY (% or s)
PRESSURE TRIGGER
FLOW TRIGGER
FLOW WAVE FORM
Obs.: Automatic Backup
(1)
1 Whenever the apnea set time is reached, the ventilator
triggers a ventilatory cycle whose configuration is based on the
current mode settings.
As soon as all ventilation parameters are received by the ventilator, it calculates the T
INS
, T
EXP
, T
PAUSE
and Ratio I:E, by function of the set Flow, Pause, Waveform, and Rate, thus obtaining all the control times to
the ventilation.
1. Ventilation without Inspiratory Pause, after T
INS
the ventilator cycles for expiration. The inspiratory
pressure reached is a consequence of the delivered volume, resistance and compliance of the
patient’s respiratory circuit.