MAGNAMED TECNOLOGIA MÉDICA S/A
In the absence of high-priority alarms, the medium priority alarm messages are displayed
alternately.
Table 21 Alarm Features
Alarm
Characteristics
High Priority
Medium Priority
Visual
Color
Red
Yellow
Intermittence frequency
1,6 hz
0,7 hz
Audible
Number of saved pulses
10 pulses
3 pulses
Interval between sage
5s
5s
Sound pressure range
65dBA
65dBA
Pulse Frequency
686 hz
686 hz