57-642 ECLIPSE Model 705 Guided Wave Radar Transmitter for Hygienic Industries
required range, as determined by the User Requirement
Specification (URS), the transmitter passes calibration veri-
fication and can be reinstalled on the process.
If the transmitter under test fails calibration verification, the
transmitter must be installed on the vessel and the
probe/transmitter must undergo a complete wet calibration.
After a successful wet calibration is completed, the transmit-
ter should be removed and taken to the dry calibration veri-
fication bench where a NEW Initial Baseline Verification
must be performed. The old Initial Baseline Verification is
now invalid as adjustments have been made to the transmit-
ter during wet calibration.
9.3 Sample Calibration Verification Document
If the accuracy is inside has limits “URS” (User
Requirement Specifications), the ECLIPSE passes.
NOTE: The accuracy required is the accuracy required per the
URS – NOT the standard accuracy of ECLIPSE!
Frequency of calibration verification is determined based on
experience, process conditions, and criticality of the mea-
surement.
INITIAL 2012 2013 2014
Mark1 (0 cm) 0
Mark2 (10 cm) 0
0
0
Mark8 (90 cm) 0
Mark9 (110 cm) 0 0,1 liter
Mark10 (110 cm) 25,2 liter 25,1 liter
Mark11 (120 cm) 45,7 liter 45,8 liter
Mark19 (180 cm) 475,3 liter 475 liter
Mark20 (190 cm) 500,2 liter 500,0 liter
Mark21 (200 cm) 500,2 liter 500,0 liter
Certified Miller Smith
Date 11.1.2010 15.1.2011

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Magnetrol eclipse Enhanced Model 705 and is the answer not in the manual?

Magnetrol eclipse Enhanced Model 705 Specifications

General IconGeneral
BrandMagnetrol
Modeleclipse Enhanced Model 705
CategoryTransmitter
LanguageEnglish

Related product manuals