Model WT3-201M Motorized Wire Crimp Pull Tester User’s Guide
9 SPEED
The speed may be adjusted to comply with various standards and test methods. Select Speed from the
menu and press ENTER. The display appears as follows:
Select inches per minute or millimeters per minute
Pull Speed
Sets the speed at which the test takes place.
Available settings:
0.4 – 12.0 in/min, in 0.1 increments, or 10 – 300 mm/min, in 0.5 mm increments
Return Speed
Sets the return speed. This speed applies to a manual press of the RETURN
key or the Auto Return sequence.
Available settings:
12 – 96 in/min, in 1 in/min increments, or 300 – 2,400 mm/min, in 25 mm/min
* in/min
mm/min
Pull Speed
4.0
Return Speed
96