Force Test Stand Model ESM301 / ESM301L
User’s Guide

Questions and Answers

  • V
    vadamsAug 19, 2025
    What to do if Mark-10 ESM301 displays 'GAUGE OVERLOAD' error?
    • B
      brandonperrySep 12, 2025
      If the Mark-10 Test Equipment's ESM301 displays a 'GAUGE OVERLOAD' error, it means the programmed overload percentage of the force gauge’s full scale has been reached. You should immediately reduce the force or adjust the overload settings.
  • D
    david52Aug 21, 2025
    What to do if Mark-10 Test Equipment displays 'GAUGE COMM ERROR' error?
    • J
      John CharlesAug 21, 2025
      If the Mark-10 Test Equipment's ESM301 displays a 'GAUGE COMM ERROR', it indicates that the communication settings in the force gauge are not correct, or the gauge is in a menu. Ensure that the force gauge is set to 9,600 baud and RS-232 communication, and that the 09-1162 interface cable is used.
  • T
    Tony JordanAug 23, 2025
    What to do if Mark-10 Laboratory Equipment ESM301 displays GAUGE OVERLOAD message?
    • C
      Cody MatthewsAug 25, 2025
      If the Mark-10 ESM301 displays the 'GAUGE OVERLOAD' message, it means the programmed overload percentage of the force gauge’s full scale has been reached. You should immediately reduce the force or adjust the overload settings.
  • S
    Sierra BennettAug 26, 2025
    Why Mark-10 Laboratory Equipment crosshead will move only in one direction?
    • E
      Erik TaylorAug 26, 2025
      If the Mark-10 Laboratory Equipment crosshead moves only in one direction, or not at all, possible causes include: * Improperly plugged cables. * The force value exceeding the overload limits configured in the gauge. * Incorrectly configured force gauge set points. Ensure all cables are correctly connected and that the force value is within the configured overload limits. Also, verify that the force gauge set points are properly configured.
  • E
    ebushAug 27, 2025
    How to fix GAUGE COMM ERROR on Mark-10 Laboratory Equipment ESM301?
    • J
      Jesus AndersonAug 27, 2025
      The 'GAUGE COMM ERROR' message on the Mark-10 ESM301 indicates that the stand is trying to move the crosshead but cannot communicate with the force gauge. This may be due to incorrect communication settings in the force gauge or the gauge being in a menu. Check the communication settings and ensure the gauge is not in a menu.
  • L
    Laura ThompsonAug 30, 2025
    What does CHECK GAUGE error mean on Mark-10 Laboratory Equipment ESM301?
    • S
      Sarah JonesAug 30, 2025
      The 'CHECK GAUGE' error message on the Mark-10 ESM301 means that the force gauge is either powered off, in a menu, or not connected to the stand controller. This message appears when overload protection is turned on. Ensure the force gauge is powered on, properly connected, and not in a menu.
  • R
    Roger ColeAug 31, 2025
    What to do if travel display value doesn't change on Mark-10 Laboratory Equipment when the crosshead moves?
    • C
      Cheyenne LoweryAug 31, 2025
      If the travel display value doesn't change when the crosshead moves on your Mark-10 Laboratory Equipment, the encoder cable might not be properly connected. Check to ensure the encoder cable is securely plugged in.
  • D
    Dana DouglasSep 3, 2025
    What does M5 1.7+ REQD mean on Mark-10 Laboratory Equipment ESM301?
    • W
      William SpearsSep 3, 2025
      The 'M5 1.7+ REQD' message on the Mark-10 ESM301 indicates that a Series 5 force gauge with firmware version 1.7 or later is required for the selected functions in the stand to operate correctly. Ensure you are using a compatible force gauge with the required firmware version.
  • M
    Mr. Jeffrey Knapp MDSep 5, 2025
    Why are some menu items missing on my Mark-10 Laboratory Equipment?
    • S
      swigginsSep 5, 2025
      If certain menu items are no longer displayed on your Mark-10 Laboratory Equipment, it may be because the demo mode has expired. All test stand functions are temporarily enabled for 160 hours of operation in Demo Mode. After this period, any functions not originally purchased will no longer be available, and their associated menu items will disappear.
  • C
    Christopher MartinSep 7, 2025
    Why does the crosshead on my Mark-10 ESM301 move in only one direction?
    • B
      Brandon ReynoldsSep 7, 2025
      If the crosshead of your Mark-10 Test Equipment moves only in one direction or not at all, check the following: * Ensure all cables are properly plugged in. * Ensure that the force value has not exceeded the overload limits configured in the stand. * Ensure that force gauge set points have been configured properly.

Summary

Setup and Safety

Mounting the Test Stand

Instructions for securely mounting the test stand to a work surface.

Safety and Proper Usage

Discusses safe operating practices and material handling precautions.

Connecting Power

Steps and safety checks for connecting the test stand to a power source.

Operation Basics

Operational Safety

Safety procedures and precautions to follow before and during test stand operation.

Test Feature Setup

Configurable Test Features

Lists test features, their status (standard/optional), and available settings.

Speed Settings

Auto Return Function

Cycling Function

Travel Limits

Overload Protection

Compression;Tension Overload

Preload Function

Load Holding Function

Break Detection

Break Detection Threshold

Break Detection Percentage

Programmable Button Function

Return to Default Settings

Password Protection

Basic Mode Operation

Button Control Modes

Overload Protection

Auto Return Mode

Cycling Mode

Preload Mode

Loadholding Mode

Break Detection Mode

Communication

PC Control Mode Commands

Function Activation

Troubleshooting

Common Error Messages

Lists common error messages and their possible causes and solutions.

Crosshead Movement Issues

Addresses problems with crosshead movement in one direction or not at all.

Specifications

Dimensions

Mark-10 ESM301 Specifications

General IconGeneral
BrandMark-10
ModelESM301
CategoryTest Equipment
LanguageEnglish