118
4888F-eIFU-0916
PERFORMANCE SPECIFICATIONS:
The table below shows Arms (Accuracy Root Mean Square) values measured using the DCI sensors with Masimo SET Oximetry
Technology in a clinical study.
SpO2
Arms
90-100% 1.44 (1,44) %
80-90% 2.30 (2,30) %
70-80% 1.84 (1,84) %
Overall 1.90 (1,90) %
SaO2 versus error (SpO2 – SaO2) with linear regression t and upper 95% and lower 95% limits of agreement.
70 -100%