Root Chapter 3: Operation
www.masimo.com 58 Masimo
Use the Sounds screen to control the volume level of sounds and duration of
audio pause for Root.
Option
Description
Factory
Default
Setting
Configurable Settings
Alarm
Volume
Sets the alarm volume
level.
Highest
volume
Slide towards the left to
decrease volume to
silence.
Pulse
Tone
Volume
Sets the pulse tone
volume level.
Highest
volume
Slide towards the left to
decrease volume to
silence.
Audio
Pause
Duration
Sets the length of time
that the audible alarm
remains silenced, when
Audio Pause is enabled.
See Audio Pause on
page 62.
2 minutes
1, 2, 3 minutes,
Permanent*, Permanent
with Reminder*.
If Permanent is selected,
there will be no audible
alarms, but visual alarms
will still display.
If Permanent with
Reminder is selected, a
tone will sound every
three (3) minutes as a
reminder that Permanent
is active.
*Requires user to have All Mute Enabled (Permanent Silence Duration) in the
Access Control menu. See Access Control on page 50.
Alarm Interface
Alarms can have different priority levels and come from different sources.
The following tables describe alarm behaviors in more detail.
Priority
Alarm Sound
1
Five (5) beeps: beep-beep-beep, brief pause, beep-beep
2
Three (3) beeps: beep-beep-beep