EasyManua.ls Logo

MediSono P3V - Menu

Default Icon
183 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
P3V Veterinary Digital Ultrasonic Diagnostic Imaging System User Manual
- 45 -
PW Invert
Invert the PW wave. (Up or Down)
D gain
Adjust D gain in the PW mode
D Frequency
Set the probe frequency in PW mode
Sample Volume
Set the default sample volume
PRF
Set the pulse repetition frequency in PW mode
Needle guide
Adjust needle guide in the B, B+M, PW mode menus.
Table 5-3 Additional Control Functions
These functions can be set using the Set and the Back keys.
5.5. Menu
Menus are displayed on the right of the screen. Only one menu can be activated at a time. The
types are shown as follows:
System status menu
In B mode or B/M mode, the system status menu provides information about the current imaging
mode. In 2B and 4B modes, it indicates the status and parameters of the active image. In M mode,
it indicates the status and parameters of M sweep. In the PW mode, it indicates the status and
parameters of Doppler wave and 2D image. The following are the system status menus of B
mode, B/M mode, M mode, and PW mode respectively.

Table of Contents