P3V Veterinary Digital Ultrasonic Diagnostic Imaging System User Manual
- 87 -
WARNING
1. Prior to each puncture, calibrate the needle guide line.
2. If the needle position is not the same as the needle guide line position, do not perform
the puncture.
To Display or To Hide the Needle Guide Line
Highlight Display in the needle guide menu, and press Set repeatedly to display or to hide the
needle guide line.
To Adjust the Needle Guide Line Position
Highlight Position in the needle guide menu, and press Set or Back to adjust the guide line
position horizontally.
To Select the Angle of Needle Guide Line
Needle guide line has been verified when the device is produced. The value is saved in Factory
data. But after a period of use, the needle guide line needs to be adjusted since the real value may
be changed. For instance, highlight Angle in the needle guide menu and press Set to adjust, and
the system displays the angle.
6.9.2. Calibrating Guide Line (Performing Phantom Cal)
1. Put the assembled puncture kit, put the probe in to water phantom, and perform the water
scan;
2. Adjust the position and angle of needle to comply with the guide line as shown below;
3. Select Verify to save the verified value.
To verify the needle guide line:
Move the needle guide line horizontally
Highlight Position, press Set to increase the value and press Back to decrease the value,
and the value is displayed in the menu option.
To adjust the angel of needle guide line:
Enter Angle option to adjust the angle, press Set to increase the value and press Back to
decrease the value, and the value is displayed in the menu option.
To save the verified value:
After verifying the position and the angle, highlight Verify and press Set, and the system
will save the verified value. After restarting the system, the verified value is activated.
To save the factory data: