88
Differential leakage current
■ ■ ■
Test voltage & frequency: Nominal mains supply
■ ■ ■
Differential leakage current accuracy: ±5% ±5 digits
■ ■ ■
Differential leakage current resolution: 0.01 mA
■ ■ ■
Display range: 0.00 to 19.99 mA
■ ■ ■
Reading corrected to nominal supply voltage +10%
■ ■ ■
Touch current test
■ ■ ■
Test voltage & frequency: Nominal mains supply
■ ■ ■
Touch current accuracy: ±5% ±5 digits
■ ■ ■
Touch current resolution: 0.01 mA
■ ■ ■
Display range: 0.00 to 10 mA
■ ■ ■
Reading corrected to nominal supply voltage +10%
■ ■ ■
Operational test
■ ■ ■
Test voltage & frequency: Nominal mains supply
■ ■ ■
VA Accuracy: ±5% ±10 digits (0 to 99 VA)
±5% ±50 digits (100 VA - 999 VA)
±5% ±100 digits (1000 VA - 3700 VA)
■ ■ ■
■ ■ ■
■ ■ ■
Resolution: 1 VA (0 to 3700 VA)
■ ■ ■
Display range: 0.00 to 3.99 KVA
■ ■ ■
Reading corrected to nominal supply voltage
■ ■ ■
Extension lead test
■ ■ ■
Tests performed: Bond, Insulation & Polarity
■ ■ ■
Polarity test voltage: 12 V dc
■ ■ ■
Polarity checks: Lead OK
Live neutral S/C
Live neutral reversed
Live/neutral O/C
■ ■ ■
■ ■ ■
■ ■ ■
■ ■ ■
Flash test
■
Flash test voltage: 1500 V a.c nominal for Class 1
■
3000 V a.c nominal for Class 2
■
Flash test current: < 3.5 mA short circuit @ 253 V primary supply voltage
■
Flash test breakdown current accuracy: ±5% ±5 digits
■
Flash test breakdown current resolution: 0.01 mA
■
Display range: 0.00 to 3.0 mA
■
Portable RCD test
■ ■ ■
Test voltage & frequency: Nominal mains supply
■ ■ ■
Test current accuracy: -8% to –2% (1⁄2 ⁄ x I)
+2% to +8% (1 x I, 5 x I)
■ ■ ■
■ ■ ■
Trip time accuracy: ±1% ±5 digits
■ ■ ■
Trip time resolution: 0.1 ms
■ ■ ■
Display range: 0 to 1999 ms (1⁄2 ⁄ x I)
0 to 300 ms (1 x I)
0 to 40 ms (5 x I)
■ ■ ■
■ ■ ■
■ ■ ■