5 First steps
19
Switching on the steam sterilizer
ü
The steam sterilizer is connected to the electricity supply.
ü
The feed water supply is secure. The steam sterilizer requires c. 3 L feed water for the first filling of the
steam generator.
1. Switch on the steam sterilizer at the power switch.
2. When the welcome screen appears, press
CONTINUE. The display switches to the main menu.
The feed water level is checked and pre-heated immediately after activation.
Depending on the device type, it requires a heat-up time of c. 9 - 13 minutes after being switched on. This
time is required for the pre-heating of the double-jacket steam generator.
Opening and closing the door
The steam sterilizer is fitted with a motor-driven automatic door locking mechanism with a threaded
spindle. Entry on the display is only possible when the door is closed.
Opening the door
The door is opened by pressing on the door symbol on the display.
When opening the door, comply with the following instructions, so as to ensure faultless operation of the
door locking mechanism.
u Never use force to open the door.
u Do not pull vigorously at the door to open it. The door opens automatically.
PLEASE NOTE
The door is to be left open only whilst loading and unloading the steam sterilizer.
Keeping the door closed saves energy.