7 Sterilization
25
7 Sterilization
Important information regarding routine operation
Please comply with the recommendations issued by the Robert-Koch-Institute (RKI) and the information
contained in DIN 58946-7 (Germany).
Manufacturer's recommendation for the routine operation of Class B steam sterilizers
1
When is it necessary to
make checks?
How should the checks be made?
Once per working day ▪ Visual check of the door seal and the door seal for damage.
▪ Check the operating agents (electricity, feed water and water
connection if necessary).
▪ Check the documentation media (printer paper / computer / network)
We recommend performing the steam penetration test with
MELAcontrol/MELAcontrol PRO in the Universal-Program (test
system in accordance with DIN EN 867-5).
Once a week ▪ Vacuum test
Tip: In the mornings before starting work - the steam sterilizer must
be cold and dry.
Batch-related tests With “Critical B” instruments:
▪ MELAcontrol/MELAcontrol PRO must be used as batch control with
every sterilization cycle.
With “Critical A” instruments:
▪ The process indicator (type 5 in accordance with DIN EN ISO 11140)
must be used as batch control with every sterilization cycle.
With “Critical A + B” instruments:
▪ MELAcontrol/MELAcontrol PRO must be used as batch control with
every sterilization cycle.
This simplifies the working procedure and increases security. You can
omit the daily steam penetration test with MELAcontrol/MELAcontrol PRO
(see above). The use of another test system in accordance with DIN EN
867-5 is possible. The number of the available test systems means that
MELAG is not able to provide technical support when using a different
system.
The indicator test strips used need not be stored.
PLEASE NOTE
The results of the tests must be documented.
1)
in accordance with the current recommendations from the Robert-Koch-Institut