13 Malfunctions
66
13 Malfunctions
Not all notifications on the display are malfunction messages. Warnings and malfunction messages are
issued on the display with an event number. This number serves identification purposes.
Type of the display
notification
Explanation
Messages A number of notifications are messages providing information.
Messages are not malfunction messages or warnings. They support
the operation of the steam sterilizer.
Warnings Warning messages are displayed when necessary. These contain
instructions which apply to you, the operator. Warning messages
are not malfunction messages. They help to ensure malfunction-
free operation and to recognize undesirable situations. Comply with
these warnings early in order to avoid malfunctions.
Malfunction messages Malfunction messages are issued when it is not possible to ensure
safe operation or safety of sterilization. These can appear on the
display shortly after switching on the device or during a program
run. If a fault occurs during a program run, the program will be
aborted. Aborting a program before the drying phase means that
the load is unsterile. If necessary, repack the load and repeat the
sterilization for the sterilization material affected.
WARNING
Danger of infection from early program abort
Aborting a program before the drying phase begins means that the load is unsterile. This
endangers the health of your patients and practice team.
n If necessary, repack the load and repeat the sterilization for the sterilization material
affected.
Before contacting customer services
Ensure that you have complied with all instructions relating to a warning or malfunction message issued on
the display of the steam sterilizer. The following table contains a summary of the most important events.
Should you be unable to find the relevant event in the table below, or your efforts do not redress the
problem, you can contact your nearest stockist or an authorized MELAG customer service provider. To
enable us to give the best possible service, please have your steam sterilizer serial number and a detailed
description of the fault to hand.
Messages
Incident Possible causes What you can do
M248 The vacuum test was performed, despite
residual moisture in the chamber or with
a load.
Repeat vacuum test, when steam sterilizer is cold
and empty