10 Settings
52
3. Confirm with SAVE.
Intelligent drying
In contrast to a conventional time-controlled drying procedure, with which the duration of the drying phase
is determined by the program, the duration of the intelligent drying is automatically calculated using the
residual moisture in the sterilization chamber. A number of factors play a role in this process including e.g.
the nature of the load, wrapped or unwrapped, the load quantity, the distribution of the load in the
sterilization chamber etc. Comply with the specifications in section Loading the steam sterilizer [}page
23].
Intelligent drying is activated in the delivery state. Should you wish to deactivate intelligent drying, proceed
as follows:
1. Select theSettingsDevice settings
Intelligent drying . The display switches
to the corresponding window.
2. If you wish to deactivate intelligent drying, select NO.
3. Confirm with SAVE.
Water supply
Select the corresponding setting in the display, depending on whether the feed water supply is effected
using the internal storage tank or a water treatment unit:
1. Select theSettings menu.
2. Navigate toDevice settingsWater
supply.
Ê The display switches to theWater supply
window.
3. Select INTERNAL, if the supply is to be performed via
the internal storage tank or EXTERNAL, if you have
connected a water treatment unit. NOTICE!When
connecting an external feed water supply it is
necessary to connect an external waste water
outflow as well. There is a risk that hot waste
water might run through the emergency overflow.
4. Confirm with SAVE.